Knowledge / Allied Health (Radiology / OT / Optometry / Dialysis) / X-ray imaging basics

X-ray imaging basics

Allied Health (Radiology / OT / Optometry / Dialysis) (B.Sc) · Radiology and imaging · free study notes for exams, placements and interviews

In short

X-ray imaging (radiography) uses a beam of high-energy photons produced in an X-ray tube that passes through the body, is attenuated differently by bone, soft tissue, fat and air, and is recorded on a detector to give a two-dimensional shadow image. It remains the first-line and highest-volume imaging investigation in India - chest, limb, spine, abdomen, dental and mobile ward radiographs - because it is quick, cheap and widely available. For B.Sc Radiology and Imaging Technology students it carries the largest share of marks in "Radiographic Physics", "Radiographic Techniques" and "Equipment" papers, and it is the core of practical and viva assessment. Recruiters - Apollo, Fortis, Max, Manipal, AIIMS and state medical colleges, Medall, Dr Lal PathLabs, Krsnaa and SRL diagnostics, plus equipment companies such as Siemens Healthineers, GE HealthCare, Philips, Wipro GE and Allengers - test exposure factors, positioning, image quality and quality-assurance knowledge in technical rounds.

Explained

Nature and production of X-rays

The X-ray tube and its components

Beam quality, quantity and exposure factors

Interaction with the patient and image formation

Image receptors: from film to digital

Image quality, geometry and radiographic technique

Indian regulation and practice (overview only - the detail belongs to radiation protection)

Key points

Formulas and examples

Maximum photon energy: E_max (keV) = kVp (numerically); minimum wavelength λ_min (nm) = 1.24 / kVp
Exposure/quantity: mAs = mA × time (s); Output ∝ mAs × kVp² (approximately)
Inverse square law: I₁/I₂ = (d₂)²/(d₁)²; Direct square (density maintenance): mAs₂ = mAs₁ × (d₂/d₁)²
15% rule: kVp₂ = 1.15 × kVp₁ with mAs₂ = mAs₁ / 2 (same receptor exposure)
Attenuation: I = I₀ e−μx; with HVL: I = I₀ × (1/2)x/HVL; μ = 0.693 / HVL
Heat units: HU = kVp × mA × s (single-phase); HU = 1.35 × kVp × mA × s (3-phase / high-frequency)
Effective focal spot = actual focal spot × sin θ (θ = target angle)
Grid ratio r = h / D; Grid (Bucky) factor = exposure with grid / exposure without grid
Magnification factor M = SID / SOD = image size / object size; % magnification = (M − 1) × 100
Geometric unsharpness Ug = f × OID / SOD, where f = effective focal spot size
Reciprocity: mA₁ × t₁ = mA₂ × t₂ for the same mAs

Symbols: kVp = peak kilovoltage (kV), mA = tube current, s = exposure time, SID = source-to-image distance, SOD = source-to-object distance, OID = object-to-image distance, HVL = half-value layer (mm Al), μ = linear attenuation coefficient (cm⁻¹ or mm⁻¹), I = intensity (mGy/min or similar), h and D in the same length unit.

Example 1 - Distance change (direct square law). A knee AP is taken at 100 cm SID with 10 mAs. The tube cannot be lowered and the exposure must be made at 120 cm. What mAs keeps the same receptor exposure?

Example 2 - 15% rule and reciprocity. A lumbar spine lateral uses 80 kVp, 200 mA, 0.4 s.

Example 3 - Attenuation and HVL. A 70 kVp beam has an HVL of 3 mm Al. What fraction is transmitted through 9 mm Al, and what is μ?

Example 4 - Magnification and geometric unsharpness. A chest is radiographed at SID 180 cm; a nodule lies 10 cm from the receptor. Effective focal spot = 1.2 mm.

Example 5 - Heat units and focal spot. A three-phase unit exposes at 80 kVp, 300 mA, 0.2 s. The target angle is 12° and the actual focal spot is 5 mm wide.

Factor increasedDensity/receptor signalContrastPatient doseScatter
kVpIncreases (∝ kVp² approx.)DecreasesDecreases for same signalIncreases
mAsIncreases proportionallyNo changeIncreasesNo change in fraction
SIDDecreases (1/d²)No changeEntrance dose decreasesSlight decrease
Added filtrationDecreasesSlight decreaseDecreases (skin dose)Little change
Grid ratioDecreases (needs more mAs)IncreasesIncreasesDecreases
Collimation (tighter)Decreases slightlyIncreasesDecreases (less volume)Decreases
ReceptorPrincipleResolutionDose efficiencyWorkflow
Film-screenSilver halide + phosphor screenHigh (fine screens)Lowest, no latitudeChemical processing, hard copy
CRStorage phosphor read by laserModerate (about 2.5-5 lp/mm)Better, wide latitudeCassette + reader, reusable plate
DR indirect (CsI/a-Si)Scintillator + TFT photodiodeGood (about 3-4 lp/mm)Highest DQEInstant, PACS/DICOM
DR direct (a-Se)X-ray to charge directlyGood to highHighInstant, PACS/DICOM

Where it is used

Interview questions

Common mistakes