Knowledge / Allied Health II (Cardiac / Respiratory / Anaesthesia / Emergency / Perfusion technology) / Oxygen therapy, aerosol therapy and airway clearance

Oxygen therapy, aerosol therapy and airway clearance

Allied Health II (Cardiac / Respiratory / Anaesthesia / Emergency / Perfusion technology) (B.Sc) · Respiratory therapy · free study notes for exams, placements and interviews

In short

Oxygen therapy is the controlled delivery of an inspired oxygen fraction above 21% to correct tissue hypoxia; aerosol therapy delivers drugs and water as suspended particles to the airway; airway clearance uses breathing techniques, pressure devices and suction to move retained secretions out. Together they are the three bedside skills that define a respiratory therapist's day in an ICU, emergency department, neonatal unit, pulmonology ward or home-care service. In Indian B.Sc Respiratory Therapy / Cardio-Pulmonary Technology programmes this topic carries the largest practical and viva weight, and it is the staple of technical rounds at Apollo, Fortis, Manipal, Narayana Health, Medanta, AIIMS and the ESIC and state-health services, as well as of oxygen-equipment and homecare employers. Medical oxygen is a drug in India (Drugs and Cosmetics Act, 1940 and Rules, 1945; monographs for *Oxygen* and *Oxygen 93 per cent* in the Indian Pharmacopoeia 2026) and is in the National List of Essential Medicines, so dose, documentation and device choice are clinical decisions, not comfort measures.

Explained

Why oxygen is given: hypoxia, hypoxaemia and targets

Oxygen sources, supply and Indian regulation

Delivery devices: low-flow, reservoir, high-flow

Humidification and the humidity deficit

Aerosol therapy: physics, devices and drugs

Airway clearance: mechanisms and techniques

Oxygen hazards, monitoring and weaning

Key points

Formulas and examples

Air : O₂ entrainment ratio = (100 − desired FiO₂%) / (desired FiO₂% − 21)
Total flow = O₂ flow × (1 + air parts), in L/min
Oxygen flow needed for a target FiO₂ in a blended/total flow : O₂ flow = Total flow × (FiO₂ − 0.21) / 0.79
Cylinder contents (L) = water capacity (L) × gauge pressure (bar or kgf/cm²)
Cylinder duration (min) = usable contents (L) / flow (L/min), keeping a residual reserve (~10% or 10-20 bar)
Liquid oxygen: 1 L liquid ≈ 860 L gas (0.86 m³)
Arterial oxygen content CaO₂ (mL/dL) = 1.34 × Hb (g/dL) × SaO₂ (fraction) + 0.003 × PaO₂ (mmHg)
Oxygen delivery DO₂ (mL/min) = CaO₂ × cardiac output (L/min) × 10
P/F ratio = PaO₂ (mmHg) / FiO₂ (fraction). ARDS (Berlin, with PEEP ≥ 5 cmH₂O): mild 200-300, moderate 100-200, severe ≤ 100.
ROX index = (SpO₂% / FiO₂) / respiratory rate (breaths/min) - used to judge HFNC success.
Humidity deficit (mg/L) = 44 − absolute humidity delivered (mg/L)
Relative humidity = (absolute humidity delivered / capacity at that temperature) × 100
Saturated absolute humidity: 20 °C = 17.3, 30 °C = 30.4, 37 °C ≈ 44 mg/L
Suction catheter size (Fr) ≈ (ETT internal diameter in mm − 1) × 2, and 1 Fr = 1/3 mm outer diameter

Example 1 - venturi mask flow check. A COPD patient is ordered 40% oxygen by air-entrainment mask with the jet set at 8 L/min oxygen.

Example 2 - cylinder duration. A D-type cylinder of 46.7 L water capacity reads 120 kgf/cm². Flow ordered 6 L/min; keep a 10% reserve.

Example 3 - oxygen content and the limits of SpO₂. Two patients, both SpO₂/SaO₂ 96%, PaO₂ 85 mmHg, cardiac output 5 L/min; Hb 14 g/dL and 9 g/dL.

Example 4 - P/F ratio and escalation. A patient on an NRBM, estimated FiO₂ 0.60, has PaO₂ 72 mmHg.

Example 5 - humidity deficit. A tracheostomised patient breathes 10 L/min of gas delivered at 30 °C, fully saturated.

Example 6 - suction catheter. ETT internal diameter 8.0 mm.

DeviceFlow (L/min)Approx FiO₂Fixed FiO₂?Typical use
Nasal cannula1-60.24-0.44NoMild hypoxaemia, eating, mobile patient
Simple mask5-100.35-0.60NoShort-term moderate need
Partial rebreathing mask8-120.40-0.70NoTransport, moderate-severe
Non-rebreathing mask10-150.60-0.80NoEmergency, trauma, CO poisoning
Venturi (air-entrainment) maskper jet0.24-0.50YesCOPD, precise titration
Air-entrainment nebuliser + T-piece10-150.28-1.0Yes (low FiO₂)Tracheostomy, post-extubation
HFNC20-600.21-1.0YesHypoxaemic respiratory failure, pre-oxygenation
Aerosol deviceDose coordinationLung dose issuesBest for
Jet nebuliserNone neededSlow, 6-8 L/min gas, dead volumeAcute attack, infant, very breathless
Vibrating meshNone neededEfficient, costlierVentilator/HFNC in-line
pMDI aloneCritical, slow deep breathHigh oropharyngeal impactionTrained adult
pMDI + spacerNot criticalBest cost-effectivenessChildren, elderly, acute asthma
DPIBreath-actuated, needs fast flowFails with weak effort/humidityStable adult, maintenance

Where it is used

Interview questions

Common mistakes

Related topics

Arterial blood gas analysis · Pulmonary function tests · Mechanical ventilation basics · Non-invasive ventilation and CPAP · Airway management and intubation · Pulmonary rehabilitation and cardiopulmonary resuscitation.