Knowledge / Dental Surgery / Oral and maxillofacial radiology

Oral and maxillofacial radiology

Dental Surgery (BDS) · Oral diagnosis and pathology · free study notes for exams, placements and interviews

In short

Oral and maxillofacial radiology is the dental speciality concerned with producing and interpreting images of the teeth, jaws, temporomandibular joints, sinuses and salivary glands for diagnosis and treatment planning. It covers intraoral projections (periapical, bitewing, occlusal), extraoral projections (panoramic, lateral cephalogram, Water's, PA mandible, reverse Towne's), and advanced imaging (CBCT, CT, MRI, ultrasound, radionuclide and contrast studies). For a BDS student it is examined in Oral Medicine and Radiology (third and final year), in NEET MDS and university practicals where film mounting, projection errors and interpretation of a given radiograph are routinely asked. In practice it is the daily working tool of every dentist in India — endodontic working-length films, implant CBCT planning, third-molar assessment, orthodontic cephalometrics and medicolegal age estimation. Radiation protection, the AERB licence and QA are handled in the separate radiation-safety topic; here the focus is technique, image geometry and interpretation.

Explained

Image formation, receptors and processing

Intraoral radiographic techniques

Extraoral and specialised imaging

Normal radiographic anatomy

Radiographic interpretation of disease

Quality, errors and Indian practice framework

Key points

Formulas and examples

Magnification M = image size / object size = SID / SOD, where SID = source-to-image (receptor) distance and SOD = source-to-object distance (same units, usually mm or cm).
Percentage magnification = (M − 1) × 100 %
Geometric unsharpness (penumbra) Ug = f × (OID / SOD), f = focal spot size (mm), OID = object-to-image distance, SOD = source-to-object distance.
Inverse square law: I₂ = I₁ × (d₁ / d₂)², I = intensity (mGy/min), d = distance from focal spot (cm or m).
Exposure compensation with distance: mAs₂ = mAs₁ × (d₂ / d₁)²
Half-value layer: I = I₀ × (1/2)x / HVL, x = absorber thickness, HVL = thickness halving intensity (mm Al or mm Pb).
Exposure (mAs) = tube current (mA) × exposure time (s)
Cieszynski's rule of isometry: central ray ⟂ bisector of the angle between the tooth long axis and the receptor plane → image length = object length.

Example 1 — Magnification in paralleling versus bisecting angle. Paralleling: SOD = 380 mm, OID = 20 mm → SID = 400 mm. M = 400/380 = 1.053 → 5.3 % magnification. Bisecting angle (short cone): SOD = 180 mm, OID = 20 mm → SID = 200 mm. M = 200/180 = 1.111 → 11.1 % magnification. A 22.0 mm root images as 22.0 × 1.053 = 23.2 mm (paralleling) versus 22.0 × 1.111 = 24.4 mm (bisecting). The paralleling film is about half as magnified — why endodontic working length is never measured off a short-cone film.

Example 2 — Geometric unsharpness. Focal spot f = 0.8 mm, OID = 20 mm, SOD = 380 mm → Ug = 0.8 × 20/380 = 0.042 mm. Same geometry with a short cone (SOD = 180 mm): Ug = 0.8 × 20/180 = 0.089 mm — more than double the blur. With a 0.4 mm focal spot at SOD 380 mm, Ug = 0.021 mm.

Example 3 — Inverse square law and mAs correction. A unit gives 2.0 mGy per exposure at 20 cm. At 40 cm: I₂ = 2.0 × (20/40)² = 0.50 mGy (one-quarter). At 200 cm (the operator's standing position): 2.0 × (20/200)² = 0.02 mGy. If a satisfactory film needed 7 mA × 0.4 s = 2.8 mAs at 20 cm, then at 40 cm: mAs₂ = 2.8 × (40/20)² = 11.2 mAs, i.e. 7 mA × 1.6 s.

Example 4 — Beam filtration and HVL. A 70 kVp beam has HVL = 2.0 mm Al. Adding 4.0 mm Al: I/I₀ = (1/2)4.0/2.0 = (1/2)² = 0.25, so 75 % of the intensity (mostly soft, skin-dose photons) is removed. Adding 6.0 mm Al: (1/2)³ = 0.125 → transmission 12.5 %.

Example 5 — Panoramic magnification and implant planning. An OPG with uniform magnification 1.25 shows 14.0 mm of bone above the mandibular canal. True height = 14.0/1.25 = 11.2 mm — a 10 mm implant is feasible but marginal. A CBCT cross-section with 0.3 mm isotropic voxels and 1:1 geometry measures 11.0 mm, confirming the estimate and additionally giving buccolingual width (6.5 mm), which the OPG cannot show. Lesson: never plan implant length from an unverified OPG measurement without a calibration ball-bearing.

FeatureIOPABitewingOPGCBCT
Field2–3 teeth, full rootCrowns + crestal boneBoth jaws, TMJ, sinusSelected volume, 3-D
ResolutionHighest (≈ 15–20 lp/mm)HighestLow–moderateModerate (voxel 0.075–0.4 mm)
Typical effective dose≈ 1–8 µSv≈ 1–8 µSv≈ 10–30 µSv≈ 20–600 µSv (field dependent)
Geometric accuracyGood (paralleling)GoodVariable magnificationNear 1:1, isotropic
Best forPeriapical disease, endodonticsProximal caries, bone levelScreening, trauma, third molarsImplants, impactions, TMJ bone
Main limitationSmall field, technique errorsNo apicesSuperimposition, ghost imagesDose, metal artefacts, no soft tissue
ErrorAppearanceCause
ForeshorteningRoots too shortExcessive vertical angulation
ElongationRoots too longInsufficient vertical angulation
OverlappingContacts superimposedWrong horizontal angulation
Cone cutClear curved borderBeam not centred on receptor
Reversed filmPale, herringbone patternFilm placed back-to-front
Blurred anteriors (OPG)Narrow, blurred incisorsPatient positioned too far forward

Where it is used

Interview questions

Common mistakes

Related topics

Radiation safety and AERB · Oral medicine and diagnosis · Oral pathology and oral microbiology · Clinical history taking and physical examination · Endodontics and conservative dentistry · Oral and maxillofacial surgery: impactions and trauma