Knowledge / Physiotherapy / Electrotherapy

Electrotherapy

Physiotherapy (BPT) · Treatment methods · free study notes for exams, placements and interviews

In short

Electrotherapy is the therapeutic use of electrical currents, electromagnetic fields, sound waves, light and thermal agents to relieve pain, stimulate nerve and muscle, promote tissue healing and reduce inflammation or oedema. It is a core clinical subject of the BPT curriculum (usually Electrotherapy I and II in the second and third year) with both theory and practical/viva papers, and it decides a large part of the skill assessment in clinical postings. Every physiotherapy job interview - hospital, polyclinic, sports or home-care - tests parameter selection, contraindications and electrical safety, because a wrong setting on an SWD or a stimulator can burn a patient. Practice in India is regulated by the National Commission for Allied and Healthcare Professions Act, 2021, under which physiotherapy is a recognised allied-health profession with state registration councils; equipment safety follows the CDSCO/BIS and Medical Devices Rules, 2017 framework. Employers expect a new graduate to set up, dose and document TENS, IFT, ultrasound, SWD, muscle stimulation, LASER and thermal modalities independently from day one.

Explained

Electrical basics and the physics behind the modalities

Low-frequency currents: stimulation and pain relief

Medium-frequency and electrodiagnosis

High-frequency and actinotherapy

Thermal, mechanical and newer agents

Safety, standards and Indian practice

Key points

Formulas and examples

Ohm's law: V = I × R — V in volts, I in amperes, R in ohms (tissue impedance)
Current density J = I / A — I in mA, A = effective electrode area in cm²; J in mA/cm² (keep below about 0.5-1 mA/cm² for comfort/safety with DC)
Charge per pulse Q = I × t — I in mA, t = pulse duration in ms, Q in microcoulombs (µC)
Duty cycle (%) = pulse duration ÷ (pulse duration + interpulse interval) × 100
IFT beat frequency f_beat = f₂ − f₁ — both in Hz
Ultrasound energy E = Intensity (W/cm²) × ERA (cm²) × time (s) → joules; total dose often quoted as W/cm² × minutes
Ultrasound treatment time ≈ (treatment area ÷ ERA) × minutes per ERA-sized zone (commonly 1-2 min per zone)
LASER energy density D = (Power in W × time in s) ÷ spot area in cm² → J/cm²; time (s) = D × area ÷ Power
Inverse square law (IRR/UVR/LASER at distance): I ∝ 1/d² → I₂ = I₁ × (d₁/d₂)²
Chronaxie: pulse duration at which the threshold current = 2 × rheobase
Iontophoresis dose = current (mA) × time (min) → mA·min (commonly 40-80 mA·min)
Heat energy in a thermal pack: Q = m × c × ΔT — m in kg, c in kJ/kg·K, ΔT in K

Example 1 - Current density and safety. A galvanic active electrode measures 4 cm × 5 cm and the machine reads 6 mA.

Example 2 - IFT beat frequency and sweep. Circuit 1 is fixed at 4,000 Hz; circuit 2 is set to 4,090 Hz.

Example 3 - Ultrasound dosage and time. A supraspinatus tendinopathy over an area of 12 cm² is treated with a 1 MHz head of ERA 5 cm², pulsed 1:4, intensity 1.0 W/cm², allowing 2 minutes per ERA-sized zone.

Example 4 - LASER exposure time. A GaAlAs 830 nm probe delivers 100 mW continuous over a spot of 0.25 cm²; the prescription is 4 J/cm².

Example 5 - Iontophoresis dose and traction load. (a) Dexamethasone iontophoresis is prescribed at 60 mA·min. At 4 mA tolerance, time = 60 ÷ 4 = 15 minutes. If the patient tolerates only 3 mA, time = 60 ÷ 3 = 20 minutes for the same dose. (b) A 70 kg adult needs lumbar traction at 30% body weight → 0.30 × 70 = 21 kgf; the same patient for cervical traction at 8% → 0.08 × 70 = 5.6 kgf, usually begun lower (about 3-4 kgf) and progressed.

ModalityFrequencyDepthMain effectKey contraindication
TENS2-150 HzCutaneous nerveAnalgesia (gate/opioid)Pacemaker, anterior neck, pregnancy (abdomen)
Faradic30-70 Hz, 0.1-1 msMotor pointTetanic contraction, muscle pumpDenervation (ineffective), fresh suture
Galvanic (DC)0 HzSkin/ionsIontophoresis, chemicalChemical burn risk, metal implants under electrode
IFT4 kHz carrier, 1-150 Hz beatDeep soft tissueAnalgesia, oedema, circulationPacemaker, malignancy, pregnancy
Ultrasound1 MHz / 3 MHz3-5 cm / 1-2 cmThermal + cavitation, tissue repairEye, uterus, heart, DVT, epiphysis
SWD27.12 MHz3-5 cmDeep heatMetal, moisture, pacemaker, sensory loss
LASER632.8 / 830-904 nm1-2 mm / up to ~3 cmPhotobiomodulation, healingEye exposure, malignancy
Wax / hot pack-≤ 1 cmSuperficial heat, extensibilityAcute inflammation, sensory loss
Cryotherapy-≤ 2 cmAnalgesia, reduced metabolismRaynaud's, cold urticaria, ischaemia
FeatureInnervated muscleDenervated muscle
Current of choiceFaradic-type (short pulse)Interrupted galvanic (long pulse)
Pulse duration0.1-1 ms10-300 ms
ResponseSmooth tetanicSluggish, worm-like
Chronaxie< 1 msOften > 10 ms
AccommodationPresentAbsent/reduced

Where it is used

Interview questions

Common mistakes

Related topics

Exercise therapy · Orthotics and prosthetics · Musculoskeletal and sports physiotherapy · Neurological rehabilitation · Human anatomy and neurophysiology of pain · Clinical assessment and electrodiagnosis (EMG/NCV)