Knowledge / Allied Health (Radiology / OT / Optometry / Dialysis) / Anaesthesia technology basics

Anaesthesia technology basics

Allied Health (Radiology / OT / Optometry / Dialysis) (B.Sc) · Operation theatre · free study notes for exams, placements and interviews

In short

Anaesthesia technology is the applied science of preparing, running and monitoring the equipment and drugs that keep a patient unconscious, pain-free and physiologically stable during surgery. The anaesthesia technologist assembles and checks the anaesthesia machine, prepares airway and vascular equipment, draws up drugs, assists intubation and spinal blocks, watches monitors and manages gas supplies, recovery and sterilisation. It is examined in B.Sc Operation Theatre / Anaesthesia Technology papers (anaesthesia equipment, pharmacology, monitoring, medical gases) and tested in placement rounds at Apollo, Fortis, Manipal, Narayana Health, Max, Medanta, AIIMS and state medical colleges, plus NHM and ESIC technician recruitment. Indian practice follows NABH and NABL standards, the Drugs and Cosmetics Act/Rules for drug and device handling, the NDPS Act for opioids, and the National Commission for Allied and Healthcare Professions Act, 2021, which now registers anaesthesia and OT technologists.

Explained

Types of anaesthesia and the stages of general anaesthesia

The anaesthesia machine (Boyle's apparatus) — gas path, component by component

Follow the gas in the order it travels; exams and interviews both ask it this way.

  1. Supply: pipeline gas at about 4 bar (400 kPa) from a central manifold through colour-coded NIST/DISS wall outlets, plus back-up cylinders on the machine yoke. Cylinder pressure is reduced by a pressure regulator to roughly the same working pressure, so the machine draws from the pipeline while cylinders stay in reserve.
  2. Indian cylinder colours (IS marking): oxygen black body, white shoulder; nitrous oxide blue; medical air grey body, white-and-black shoulder; carbon dioxide grey; entonox blue with white-and-blue quarters. Many hospitals also use the all-white ISO oxygen cylinder — read the label, never the paint alone.
  3. Pin Index Safety System (PISS) for small cylinders, non-interchangeable screw threads (NIST) for pipelines, and colour coding — three independent safeguards against connecting the wrong gas.
  4. Flowmeters (rotameters): tapered glass tubes with a rotating bobbin; one per gas, reading in L/min. Oxygen rotameter is placed downstream (nearest the common outlet) so a leak in another tube cannot rob the mixture of oxygen.
  5. Safety interlocks: oxygen failure (Ritchie/Bosun) alarm sounds below ~200 kPa; oxygen-nitrous oxide ratio device (hypoxic guard) limits N₂O so FiO₂ cannot fall below about 25%; fail-safe valve cuts N₂O when oxygen pressure falls; non-return valve prevents backflow into the pipeline.
  6. Vaporiser: adds a measured percentage of volatile agent. Modern agent-specific, temperature- and flow-compensated, variable-bypass, out-of-circuit (VOC) vaporisers — Tec series for halothane, isoflurane, sevoflurane; desflurane needs an electrically heated, pressurised vaporiser (boiling point about 23 °C). Selectatec back-bar mounting with interlock so only one vaporiser can be switched on.
  7. Common gas outlet → breathing system. Oxygen flush bypasses the vaporiser and delivers 35-75 L/min of pure oxygen — never press it during spontaneous ventilation on a Bain circuit or with a patient connected to a ventilator (barotrauma risk, and it dilutes the agent).
  8. Scavenging system: active (vacuum) or passive, to remove waste anaesthetic gases from theatre air.
  9. Ventilator and bellows: ascending (standing) bellows are preferred — they collapse visibly on a disconnection. Modes: volume-controlled, pressure-controlled, SIMV, pressure support.

Breathing systems and the circle absorber

Airway equipment and intubation

Drugs the technologist must know and prepare

Monitoring, safety checks and medical gas systems

Complications, recovery and legal-professional framework

Key points

Formulas and examples

Minute ventilation MV = Tidal volume (V_T) × Respiratory rate (RR). V_T in mL (6-8 mL/kg ideal body weight), RR per min, MV in mL/min or L/min.
Paediatric uncuffed ETT internal diameter (mm) = (Age in years / 4) + 4; oral insertion depth (cm) ≈ 3 × ID.
Cylinder contents: for a gas-filled cylinder (O₂), Volume at 1 atm ≈ Gauge pressure (bar) × Water capacity (L). Duration (min) = Contents (L) / Flow (L/min).
Total MAC = Σ (delivered % of each agent / its MAC%); 1.0 MAC-equivalent is the usual surgical target with opioids.
Maximum safe local anaesthetic dose (mg) = Allowed mg/kg × Body weight (kg); volume (mL) = Dose (mg) / (10 × % concentration), since a 1% solution = 10 mg/mL.
Fresh gas flow for a circle at steady state ≈ O₂ consumption (≈ 3-4 mL/kg/min) in a fully closed system; low-flow practice uses 0.5-1 L/min after an initial high-flow wash-in.
Alveolar gas equation (examined in physiology-linked papers): PAO₂ = FiO₂ × (Patm − PH₂O) − PaCO₂ / R, with Patm = 760 mmHg, PH₂O = 47 mmHg, R ≈ 0.8.

Example 1 - Ventilator setting. A 70 kg adult for laparotomy, lung-protective setting at 7 mL/kg, RR 12/min.

Example 2 - Oxygen cylinder duration. A size-D (water capacity 2.3 L) emergency oxygen cylinder for transport reads 120 bar on the gauge; the patient needs 6 L/min.

Example 3 - Local anaesthetic maximum dose. A 60 kg woman for an axillary block with lignocaine plus adrenaline (7 mg/kg).

Example 4 - Paediatric airway and tube. An 8-year-old child, 24 kg, elective hernia repair.

Example 5 - MAC arithmetic. Maintenance with 60% nitrous oxide and 0.8% isoflurane in oxygen.

FeatureSuccinylcholineRocuroniumAtracurium
ClassDepolarisingNon-depolarising (steroidal)Non-depolarising (benzylisoquinolinium)
Intubating dose1-1.5 mg/kg0.6-1.2 mg/kg0.5 mg/kg
Onset30-60 s60-90 s (high dose)2-3 min
Duration5-10 min30-60 min20-35 min
EliminationPlasma cholinesteraseLiver/bileHofmann + ester hydrolysis
ReversalNone neededNeostigmine or sugammadexNeostigmine
Key cautionHyperkalaemia, MH, bradycardia in childrenAnaphylaxis, prolonged in liver diseaseHistamine release, keep refrigerated
FeatureSpinal (subarachnoid)Epidural
SpaceSubarachnoid, below L2Epidural, any level
Needle25-27 G pencil-point16-18 G Tuohy, catheter
Drug volume2-3 mL heavy bupivacaine10-20 mL, top-ups possible
Onset2-5 min15-20 min
Block qualityDense, reliableSegmental, titratable
Main risksHypotension, PDPH, total spinalPatchy block, dural puncture, epidural haematoma

Where it is used

Interview questions

Common mistakes

Related topics

Operation theatre technology · Human anatomy and physiology (respiratory and cardiovascular) · Medical gas pipeline systems and cylinder handling · Patient monitoring instrumentation · Sterilisation, disinfection and biomedical waste management · Critical care and ventilator basics