Bioelectric signals and electrodes

Biomedical Engineering (B.E./B.Tech) · Biomedical instrumentation and signals · free study notes for exams, placements and interviews

In short

A bioelectric signal is the voltage that appears outside living cells because excitable membranes pump ions unequally and then let them move: the resting membrane potential and the action potential are its source, and ECG, EEG, EMG, EOG, ERG and EGG are its measurable surface forms. An electrode is the transducer that converts this ionic current in tissue into an electronic current in a wire, so its half-cell potential, impedance and artifact behaviour decide whether a 50 µV EEG wave survives at all. This topic is the foundation of every biomedical instrumentation paper, carries direct GATE (BM) marks on the Nernst/Goldman equations, electrode models and signal specifications, and is asked in interviews at BPL Medical Technologies, Schiller Healthcare India, Allengers, Skanray, Wipro GE, Philips, Siemens Healthineers, Nihon Kohden India and hospital biomedical-engineering departments. In practice it governs electrode selection, skin preparation, lead placement and the safety limits written into IEC 60601 as adopted by BIS and enforced through the Medical Devices Rules, 2017.

Explained

Origin of bioelectric potentials

The bioelectric signals and their specifications

Electrode-electrolyte interface and half-cell potential

Types of electrodes and their classification

  1. Surface (macro) electrodes - record from the skin.
    • Metal plate: German-silver or nickel-plated limb plates with rubber straps and electrolyte paste; cheap, reusable, prone to motion artifact.
    • Suction cup (Welch) electrode: precordial ECG, no strap needed.
    • Floating / recessed electrode: the Ag/AgCl disc never touches skin; a gel-filled cavity isolates it, so mechanical disturbance of the skin does not disturb the metal-electrolyte interface - the standard anti-motion-artifact design.
    • Disposable pre-gelled foam/tape electrodes with sintered Ag/AgCl and solid hydrogel - the ward and ICU default; snap or tab connection.
    • Flexible / textile / conductive-polymer electrodes for neonates, long-term Holter and wearables.
    • Dry and capacitive (insulated) electrodes: no gel; capacitive coupling through an insulator with a very high input-impedance buffer. Convenient for wearables but noisy and motion-sensitive.
    • Ear-clip, scalp-cup (collodion/paste-fixed), bridge and spray electrodes for EEG and ERG.
  2. Needle and wire (intramuscular) electrodes - EMG and nerve conduction studies.
    • Monopolar needle (insulated except the tip, against a surface reference), concentric (coaxial) needle - the clinical standard, picks up a small territory, bipolar concentric (two wires in one cannula, most selective), fine-wire hooked electrodes for kinesiology.
  3. Microelectrodes - tip 0.5-5 µm, record a single cell's membrane potential.
    • Glass micropipette filled with 3 M KCl, with an Ag/AgCl wire; resistance 1-100 MΩ, tip potential and high noise are its drawbacks.
    • Metal microelectrodes: electrolytically etched tungsten or stainless steel, insulated to the tip; mainly capacitive impedance.
    • Solid-state / MEMS arrays: silicon multi-site probes, Utah and Michigan arrays, microelectrode arrays (MEA) for in-vitro cardiac and neural work.
  4. Stimulating electrodes (pacing leads, defibrillator pads, TENS, deep-brain stimulation) - must be charge-balanced and inert (Pt, Pt-Ir, Ir oxide, titanium nitride) to avoid tissue damage and electrode dissolution.
  5. Internal / indwelling - catheter-tip, oesophageal, intracardiac (EP mapping), cortical strip/grid for electrocorticography.

Placement, lead systems and reference schemes

Noise, artifacts and remedies

Standards, safety and Indian practice

Key points

Formulas and examples

Nernst: E = (RT / zF) · ln(C_out / C_in) volts
R = 8.314 J mol⁻¹ K⁻¹, T = absolute temperature (K), F = 96,485 C/mol, z = valence (signed), C = ionic activity ≈ concentration (mmol/L). At 37 °C, RT/F = 26.7 mV; with log₁₀ the factor is 61.5 mV/decade for z = 1.
Goldman-Hodgkin-Katz: V_m = (RT/F) · ln{ (P_K[K]_o + P_Na[Na]_o + P_Cl[Cl]_i) / (P_K[K]_i + P_Na[Na]_i + P_Cl[Cl]_o) }
P = membrane permeability (m/s); note chloride is inverted because z = −1.
Electrode impedance: Z(jω) = R_s + R_d / (1 + jωR_d C_d); |Z| → R_s + R_d at ω → 0, → R_s at ω → ∞; corner f_c = 1/(2πR_d C_d) Hz
Common-mode to differential conversion by imbalance: V_diff ≈ V_cm · ΔZ_e / Z_in, where ΔZ_e = electrode impedance mismatch (Ω) and Z_in = amplifier input impedance per lead (Ω)
CMRR: CMRR(dB) = 20 log₁₀(A_d / A_cm); output error referred to input = V_cm / 10CMRR/20
Thermal (Johnson) noise of electrode resistance: V_n = √(4 k T R B) volts rms; k = 1.38 × 10⁻²³ J/K, B = bandwidth (Hz)
Charge injection (stimulation): Q = I × t coulombs; charge density = Q / A (µC/cm²), kept below the electrode material's safe limit

Example 1 - Nernst potential for potassium. [K]_in = 140, [K]_out = 4 mmol/L, T = 310 K, z = +1.

Example 2 - resting potential from the Goldman equation. P_K : P_Na : P_Cl = 1 : 0.04 : 0.45, same concentrations.

Example 3 - electrode impedance against frequency. R_s = 1 kΩ, R_d = 50 kΩ, C_d = 100 nF.

Example 4 - why unequal electrodes put 50 Hz on one ECG lead. V_cm = 2 V at 50 Hz on the body, Z_in = 10 MΩ per input, electrode impedances 10 kΩ and 30 kΩ (ΔZ = 20 kΩ).

Example 5 - amplifier CMRR and microelectrode noise. (a) CMRR = 100 dB, V_cm = 1.5 V: error = 1.5/10⁵ = 15 µV, i.e. 1.5% of a 1 mV ECG but 30% of a 50 µV EEG wave - EEG front ends need ≥ 110-120 dB.

PropertyAg/AgCl (non-polarizable)Platinum / stainless steel (polarizable)
Charge transferFaradaic, reversibleMainly capacitive
Impedance at low fLow and nearly flatVery high
DC stability / driftExcellentPoor
NoiseLowHigher
Best useECG, EEG, EMG, EOG recordingStimulation, implants, harsh environments
ElectrodeTypical sizeImpedance (10 Hz)Main limitation
Metal plate, paste30-50 mm2-10 kΩMotion artifact
Disposable pre-gelled Ag/AgCl10-20 mm1-5 kΩSingle use, skin irritation
Floating / recessed10-20 mm1-5 kΩCost, bulk
EEG cup6-10 mm5-20 kΩFixing time, hair
Concentric needle0.3-0.5 mm10-100 kΩInvasive, sterilisation
Glass micropipette0.5-5 µm tip1-100 MΩFragile, very noisy

Where it is used

Interview questions

Common mistakes

Related topics

Biomedical instrumentation · ECG, EEG and EMG instrumentation · Signal conditioning and data acquisition · Biomedical signal processing · Patient monitoring and life-support equipment · Electrical safety and medical device standards.