Knowledge / Biomedical Engineering / ECG, EEG and EMG instrumentation

ECG, EEG and EMG instrumentation

Biomedical Engineering (B.E./B.Tech) · Biomedical instrumentation and signals · free study notes for exams, placements and interviews

In short

ECG, EEG and EMG instrumentation is the chain of electrodes, protection circuits, high-CMRR amplifiers, filters, isolation and digitisers that turns millivolt- to microvolt-level bioelectric potentials from the heart, brain and muscle into a trace or a data file. The three machines share the same architecture but differ hugely in amplitude, bandwidth and gain, which is exactly what examiners test. The topic carries steady marks in university Biomedical Instrumentation papers and appears in GATE BM (biomedical signal acquisition, amplifier specifications), and in interviews at Indian employers such as BPL Medical Technologies, Schiller Healthcare India, Allengers, Recorders & Medicare Systems (RMS), Nihon Kohden India, Philips/GE Healthcare Bengaluru, SCTIMST and AIIMS/PGI biomedical engineering departments. Field service, R&D, test-and-validation and hospital biomedical-engineering roles all need it, along with the safety standards (IEC 60601 series, adopted in India as IS 13450 parts and enforced through CDSCO licensing of medical devices).

Explained

Amplitude, bandwidth and gain - the three machines compared

The common front end: electrodes, protection, amplifier

  1. Electrodes: Ag/AgCl non-polarisable electrodes with conductive gel. ECG - limb plates/suction and disposable pre-gelled Ag/AgCl discs; EEG - cup/disc electrodes with collodion or paste, or a cap; EMG - surface bar electrodes, concentric needle, monopolar needle, single-fibre and fine-wire. Skin preparation to bring impedance below ~5 kΩ (EEG below 5 kΩ, ideally balanced between electrodes).
  2. Defibrillation and ESU protection: series current-limiting resistors (~10 kΩ-1 MΩ), neon lamp/gas discharge tube or back-to-back diode clamps, spark gaps; protects the input from a 5 kV defibrillator discharge. Required by IEC 60601-2-25 (ECG) and -2-27 (monitors).
  3. Lead-off / electrode-contact detection: a small high-frequency or d.c. current injected to flag a disconnected electrode.
  4. Instrumentation amplifier: three-op-amp IA (or a dedicated chip - AD620/INA118/AD8232 ECG front end, or an integrated ADS1298 8-channel 24-bit AFE used in modern ECG boards). High input impedance (>10 MΩ, often >100 MΩ), CMRR > 100 dB (ECG ≥ 89 dB per standard; EEG 100-120 dB), low input noise (<1 µV p-p for EEG), input bias current in nA/pA.
  5. Right-leg drive (driven right leg, DRL): the right-leg electrode is not grounded; the sensed common-mode voltage is inverted, amplified and fed back to the body, cutting effective common-mode interference by 10-40 dB while limiting fault current. Modern standard practice.
  6. Filters: a.c.-coupling high-pass (0.05 Hz ECG; 0.5 Hz EEG) to block electrode d.c. offset; anti-aliasing low-pass before the ADC; a 50 Hz notch (India's mains frequency - 60 Hz in the Americas), used sparingly in diagnostic ECG because it distorts the QRS.
  7. Isolation: optical or transformer-coupled isolation amplifier plus isolated d.c.-d.c. supply, giving a Type CF applied part (floating, defibrillation-proof, patient auxiliary current ≤10 µA normal / 50 µA single fault for CF). Protects against macro- and micro-shock.
  8. Output: ADC (12-24 bit, 500-1,000 samples/s per channel for ECG; 250-512 for EEG; 2-10 kHz for EMG), DSP, display, thermal-array writer.

ECG instrumentation: leads and the recorder

EEG instrumentation

EMG instrumentation and nerve conduction

Safety, standards and Indian regulation

Key points

Formulas and examples

Differential amplifier output: Vo = Ad(V1 − V2) + Acm·(V1 + V2)/2, where Ad = differential gain, Acm = common-mode gain (dimensionless)
CMRR = 20 log10 (Ad / Acm) in dB
Three-op-amp instrumentation amplifier gain: Ad = (1 + 2R1/RG) × (R3/R2); for AD620-type, G = 1 + 49.4 kΩ/RG
High-pass (a.c. coupling) cut-off: fc = 1/(2πRC) Hz, R in Ω, C in F
Notch/bandstop centre: f0 = 1/(2πRC) for a twin-T; quality factor Q = f0/BW
Nyquist: fs ≥ 2 fmax; ADC resolution q = VFS / 2n volts per LSB (n = bits)
SNR after N-sweep averaging: SNRN = √N × SNR1 (amplitude SNR)
Heart rate from ECG paper at 25 mm/s: HR (bpm) = 1500 / (R-R in mm) = 60 / (R-R in s)
Nerve conduction velocity: NCV = distance between stimulation sites / (proximal latency − distal latency), m/s
EMG RMS over window T: Vrms = √[(1/T)∫ v²(t) dt], volts
Interference voltage from displacement current: Vcm ≈ Icm × Zbody-to-earth; differential error ≈ Icm × ΔZelectrode

Example 1 - CMRR requirement for ECG. A patient picks up a 50 Hz common-mode voltage of 2 V. The ECG signal is 1 mV and the allowable 50 Hz artefact at the output, referred to input, is 10 µV (1% of signal).

Example 2 - ECG heart rate and interval reading. A lead II trace at 25 mm/s, 10 mm/mV shows R-R = 20 mm, QRS width 2.5 mm, R-wave height 13 mm.

Example 3 - Instrumentation amplifier gain and a.c. coupling for EEG. Design for a 50 µV peak EEG signal to give ±2.5 V at the ADC input, with a 0.5 Hz high-pass.

Example 4 - Evoked potential averaging. A visual evoked potential of 5 µV is buried in 50 µV of background EEG (SNR = 0.1, i.e. −20 dB).

Example 5 - Nerve conduction velocity and sampling. Median nerve stimulated at the wrist (distal latency 3.4 ms) and the elbow (proximal latency 7.6 ms); the two sites are 24 cm apart. EMG recorded at 10 kHz, 12-bit ADC, ±5 V full scale.

FeatureECGEEGEMG
SourceCardiac muscle depolarisationCortical post-synaptic potentialsMuscle fibre action potentials
Amplitude0.5-5 mV10-100 µV50 µV-10 mV
Bandwidth0.05-150 Hz0.5-70 Hz10 Hz-10 kHz
Typical gain~1,00010,000-100,0001,000-10,000
ElectrodesAg/AgCl discs, limb platesCup/disc, cap (10-20)Needle, surface bar, fine wire
Sampling rate500-1,000 Hz250-512 Hz2-40 kHz
Channels3, 6, 12 (12-lead)8-64+1-4 (plus stimulator)
Display standard25 mm/s, 10 mm/mV30 mm/s, ~7 µV/mmOscilloscope + audio
Key hardware needDefib protection, baseline restorerVery low noise, montage selectorWide band, isolated stimulator
Filter in the chainPurposeTypical setting
High-pass (a.c. coupling)Blocks ±300 mV electrode d.c. offset, cuts baseline wander0.05 Hz ECG diagnostic; 0.5 Hz EEG; 10-20 Hz EMG
Low-pass (anti-alias)Prevents aliasing, removes EMG/RF noise150 Hz ECG; 70 Hz EEG; 5-10 kHz EMG
Notch (bandstop)Removes 50 Hz mains in India50 Hz, narrow Q; avoid in diagnostic ECG
Adaptive/DSPBaseline correction, motion-artefact removalWavelet, adaptive LMS in digital machines

Where it is used

Interview questions

Common mistakes

Related topics

Bioelectric signals and electrodes · Signal conditioning and data acquisition · Biomedical signal processing · Digital signal processing: filters and FFT · Patient monitoring and life-support equipment · Medical device safety standards and electrical safety testing