Electronic instruments: CRO and multimeter

Electronics (B.Sc) · Practice and careers · free study notes for exams, placements and interviews

In short

A cathode ray oscilloscope (CRO) draws a graph of voltage against time on a screen, so it shows *shape* - amplitude, period, rise time, distortion, phase and noise - while a multimeter gives a single steady number for voltage, current, resistance or continuity. Between them they handle almost every bench measurement in electronics, which is why both appear in every B.Sc practical syllabus, in the "Electronic Measurements and Instrumentation" paper of UGC-CBCS and AICTE curricula, and in GATE EC/IN and PSU questions on deflection sensitivity, dual-slope ADCs, loading error and probe loading. Interviewers at test-and-measurement, service and EMS companies (Keysight and Tektronix channel partners, Scientific/Aplab, Meco, HTC, Rishabh, Fluke India, plus Bharat Electronics, ISRO, DRDO labs and every electronics service centre) open with "how will you measure this?" and judge the answer on instrument choice, range, loading and safety. Calibration practice in India runs through NABL-accredited labs with traceability to CSIR-NPL, New Delhi, and safety through the IEC 61010 series - IS/IEC 61010-1 for the instrument, IEC 61010-2-030 for the measurement categories (CAT II-IV) of its inputs.

Explained

The cathode ray tube and electrostatic deflection

Block diagram and the time base

Multi-trace, Lissajous figures and probes

Analog versus digital storage oscilloscope (DSO)

The analog multimeter (VOM)

The digital multimeter (DMM)

Key points

Formulas and examples

CRT deflection: y = (L · l · V_d) / (2 · d · V_a) — y = deflection on screen (m), L = distance from plate centre to screen (m), l = plate length (m), d = plate separation (m), V_d = deflecting voltage (V), V_a = accelerating voltage (V)
Deflection sensitivity S = y / V_d = (L · l) / (2 · d · V_a) in m/V; deflection factor G = 1/S in V/m
Peak-to-peak voltage V_pp = (vertical divisions) × (volts/div) × (probe factor); V_rms = V_pp / (2√2) for a sine wave
Time period T = (horizontal divisions) × (time/div); frequency f = 1/T; phase shift φ° = (Δx divisions / T divisions) × 360°
Lissajous: f_vertical / f_horizontal = (tangencies to a horizontal line) / (tangencies to a vertical line); sin φ = y₁ / y₂ (y₁ = y-intercept, y₂ = maximum y-deflection)
Scope rise time t_r(scope) ≈ 0.35 / BW; measured t_r = √(t_r,signal² + t_r,scope²)
Nyquist: f_s ≥ 2 f_max; record length = memory depth / sample rate; DSO resolution = full-scale span / 2ⁿ (n = ADC bits)
Analog meter: sensitivity = 1 / I_fsd (Ω/V); multiplier R_s = (V_range / I_fsd) − R_m; shunt R_sh = (I_m · R_m) / (I − I_m); voltmeter input resistance R_in = sensitivity × range
DMM: resolution = range / counts; error = ±(a% of reading + b counts); RMS value V_rms = √(mean of v²)

Example 1 - Reading a waveform off the screen. A sine wave with a ×10 probe occupies 6.0 vertical divisions peak-to-peak at 0.5 V/div, and one cycle spans 4.0 horizontal divisions at 0.2 ms/div.

Example 2 - Deflection sensitivity of a CRT. Plate length l = 2 cm, separation d = 0.5 cm, plate-centre-to-screen L = 20 cm, accelerating voltage V_a = 2,000 V.

Example 3 - Analog voltmeter design and its loading error. A 50 µA, 2 kΩ PMMC movement is used on a 10 V DC range.

Example 4 - Bandwidth and rise time. A 20 MHz CRO measures a pulse edge as 40 ns.

Example 5 - DMM resolution and accuracy. A 3½-digit DMM (2000 counts, reading up to 1999) specified ±(0.5% of reading + 2 counts) reads 1.234 V on the 2 V range.

Example 6 - Lissajous phase and frequency. An ellipse has y-intercept y₁ = 2.0 div and maximum vertical deflection y₂ = 4.0 div → sin φ = 0.5 → φ = 30°. A pattern with 3 tangencies to a horizontal line and 2 to a vertical line gives f_v/f_h = 3/2, so a 1 kHz horizontal input implies 1.5 kHz on the vertical.

FeatureAnalog CRODSOAnalog multimeterDMM
OutputContinuous traceSampled, stored tracePointer on scaleDigits
Shows waveform shapeYesYesNoNo
Single-shot captureNoYes (with pre-trigger)NoNo
Typical input resistance (DC V)1 MΩ (10 MΩ with ×10 probe)1 MΩ20 kΩ/V × range10 MΩ
Reads non-sine AC correctlyYes (shape visible)YesNo (sine-calibrated)Only if true-RMS
Best forCheap teaching lab, no aliasingGlitches, digital timing, loggingNulling, trend watchingAccurate numeric readings
WeaknessNo storage, manual readingAliasing, 8-bit resolutionLoading, parallax, class 1-2 errorNo waveform information

Where it is used

Interview questions

Common mistakes

Related topics

Measurement errors, accuracy and calibration · Bridges and transducers for R, L, C measurement · Function generators, frequency counters and signal sources · Rectifiers, filters and regulated power supplies · Digital-to-analog and analog-to-digital converters · Electronics careers: service, testing and embedded