Pressure measurement

Instrumentation and Control Engineering (B.E./B.Tech) · Sensors and measurement · free study notes for exams, placements and interviews

In short

Pressure measurement is the determination of normal force per unit area exerted by a fluid, either against vacuum (absolute), against the local atmosphere (gauge), or between two points (differential). It is the most widely measured process variable in industry after temperature, and because differential pressure is also used to infer flow and level, a pressure element sits behind a very large share of all plant instrumentation. Indian university syllabi and GATE (Instrumentation) examine manometry, elastic elements, electrical pressure transducers, vacuum gauges and dead-weight calibration; interview rounds at Honeywell, Emerson, Yokogawa, ABB, Siemens, L&T, BHEL, Forbes Marshall, Reliance, IOCL and NTPC go straight to Bourdon tubes, DP transmitters, 4-20 mA loops, impulse-line errors and calibration traceability. Legal and safety practice in India comes through the Indian Boiler Regulations, PESO rules for pressure vessels and gas cylinders, the Legal Metrology Act, 2009 for trade measurement, and NABL/CSIR-NPL traceability for calibration.

Explained

Definitions, references and units

Liquid-column (manometric) methods - primary standards for low pressure

Elastic (mechanical) elements

Electrical and electronic pressure transducers

Vacuum and very high pressure

Installation, errors and dynamic behaviour

Key points

Formulas and examples

p = F/A [Pa = N/m²]; p = ρ g h [ρ = density kg/m³, g = 9.81 m/s², h = head m]
U-tube differential: p₁ − p₂ = (ρ_m − ρ_f) g h [ρ_m = manometer liquid, ρ_f = fluid above it]
Inclined manometer: p = ρ g L sinθ; sensitivity = L/h = 1/sinθ
Two-liquid micromanometer: Δp = (ρ₁ − ρ₂) g h (reservoir-area correction neglected)
Absolute = gauge + atmospheric; vacuum (abs) = p_atm − vacuum reading
Flat clamped circular diaphragm, centre deflection y = 3 p R⁴ (1 − ν²) / (16 E t³) [R = radius, t = thickness, E = Young's modulus, ν = Poisson's ratio]; keep y < t/3 for linearity
Strain-gauge full bridge: V_o = V_ex · GF · ε (four active gauges); quarter bridge: V_o = V_ex · GF · ε / 4
Capacitive sensor: C = ε₀ε_r A / d; differential output ∝ (C₁ − C₂)/(C₁ + C₂)
Piezoelectric: Q = d·F = d·p·A; open-circuit voltage V = Q/C
Dead-weight tester: p = M g / A_eff
McLeod gauge: p = a h² / (V − a h) ≈ a h² / V
Current-loop conversion: p = (I − 4 mA)/(16 mA) × span + LRV; V across R = I·R
Unit conversions: 1 atm = 101.325 kPa = 760 mmHg = 1.01325 bar = 10.33 mH₂O; 1 torr = 133.322 Pa; 1 kgf/cm² = 98.0665 kPa; 1 mmWC = 9.81 Pa

Example 1 - mercury manometer, gauge and absolute. A U-tube mercury manometer connected to a gas line reads h = 250 mm. ρ_Hg = 13,600 kg/m³, g = 9.81 m/s², barometer 101.325 kPa.

Example 2 - inclined draft gauge. Tube at 30° to horizontal, liquid SG 0.8, reading along the tube L = 150 mm.

Example 3 - dead-weight tester. Piston effective area A_eff = 1.000 cm² = 1.000 × 10⁻⁴ m²; total mass loaded (piston + weights) = 20.00 kg; local g = 9.790 m/s².

Example 4 - diaphragm sizing. Stainless-steel flat diaphragm, R = 10 mm, t = 0.3 mm, E = 200 GPa, ν = 0.3, p = 1 bar = 10⁵ Pa.

Example 5 - 4-20 mA transmitter and piezoelectric sensor.

Example 6 - vacuum conversions. A gauge shows 600 mmHg vacuum with barometer 760 mmHg.

DeviceTypical rangeAccuracy (typical)Notes
U-tube / inclined manometer0-5 kPa (to ~2 bar with Hg)0.1-1%Primary, no power, bulky
Bourdon dial gauge0.6 bar to 700 bar0.5-2.5% of spanLocal indication, hysteresis
Diaphragm / capsule0-0.1 bar to 10 bar0.5-1%Low pressure, corrosive service
Bellows0-0.05 to 5 bar0.5-1%Switches, DP units
Strain gauge / piezoresistive0-0.1 bar to 1000 bar0.1-0.5%Small, fast, temp-compensated
Capacitive DP transmitter0-0.1 kPa to 20 MPa0.05-0.1% of spanHigh turndown, HART
Piezoelectricto 10 kbar, dynamic only1%kHz response, no static reading
Pirani gauge10⁻⁴ to 1 mbar10-20%Gas dependent
Ionization gauge10⁻³ to 10⁻¹⁰ mbar10-30%High vacuum only
Dead-weight tester0.2 bar to 1000+ bar0.015-0.05%Calibration standard

Where it is used

Interview questions

Common mistakes

Related topics

Transducers and sensors basics · Flow measurement · Level measurement · Temperature measurement · Signal conditioning and bridge circuits · Process control and final control elements