In short
Pressure measurement is the determination of normal force per unit area exerted by a fluid, either against vacuum (absolute), against the local atmosphere (gauge), or between two points (differential). It is the most widely measured process variable in industry after temperature, and because differential pressure is also used to infer flow and level, a pressure element sits behind a very large share of all plant instrumentation. Indian university syllabi and GATE (Instrumentation) examine manometry, elastic elements, electrical pressure transducers, vacuum gauges and dead-weight calibration; interview rounds at Honeywell, Emerson, Yokogawa, ABB, Siemens, L&T, BHEL, Forbes Marshall, Reliance, IOCL and NTPC go straight to Bourdon tubes, DP transmitters, 4-20 mA loops, impulse-line errors and calibration traceability. Legal and safety practice in India comes through the Indian Boiler Regulations, PESO rules for pressure vessels and gas cylinders, the Legal Metrology Act, 2009 for trade measurement, and NABL/CSIR-NPL traceability for calibration.
Explained
Definitions, references and units
- Pressure p = F/A, normal force per unit area; SI unit pascal (Pa = N/m²). Practical units: kPa, MPa, bar (10⁵ Pa), kgf/cm² (98.0665 kPa, still printed on Indian gauges), psi, mmHg (torr), mmWC.
- Absolute pressure: measured from perfect vacuum (zero). Used for vacuum work, gas-law calculations and compressor suction. Written bar(a) or kPa(a).
- Gauge pressure: measured above local atmospheric; most industrial gauges read this, because the case interior is at atmosphere. bar(g). Absolute = gauge + atmospheric.
- Vacuum: negative gauge pressure, usually quoted as mmHg vacuum below atmosphere.
- Differential pressure (DP): difference between two unknown pressures; the basis of orifice flow, filter clogging and level in closed vessels.
- Static vs dynamic: static pressure acts on a surface parallel to flow; total (stagnation) pressure = static + dynamic (½ρV²), measured by a Pitot tube.
- Head: pressure expressed as a height of liquid column, p = ρgh - the link between manometry and pressure.
- Terms graded in exams: range, span (upper minus lower range value), turndown (max span/min calibrated span), accuracy (% of span), linearity, hysteresis, repeatability, resolution, dead band, drift, over-range limit, burst pressure.
Liquid-column (manometric) methods - primary standards for low pressure
- U-tube manometer: p₁ − p₂ = ρ g h. Simple, self-calibrating (needs only density and a length), but limited to roughly a few bar before the column becomes impractical; two menisci to read.
- Well-type (cistern) manometer: one large-area reservoir, one tube; only one reading needed, with a small correction for well level fall (a/A term).
- Inclined-tube manometer: tube at angle θ, so measured length L = h/sinθ. Sensitivity multiplier = 1/sinθ (θ = 30° doubles, 10° gives ≈ 5.76×). Used as draft gauges for furnace and duct pressures in mmWC.
- Two-liquid (differential) micromanometer: two immiscible liquids of close densities ρ₁, ρ₂ in enlarged reservoirs; effective density becomes (ρ₁ − ρ₂), giving very large magnification for pressures of a few pascals.
- Ring-balance and bell-type manometers: a sealed ring or bell floats in liquid and rotates/lifts against a counterweight, giving large torque at very low DP.
- Barometers: Fortin mercury barometer (primary atmospheric standard, with temperature, capillarity and gravity corrections) and the aneroid barometer (evacuated capsule).
- Corrections: liquid density at actual temperature, local g, capillary depression of mercury (use wide bore), zero error, parallax, tube verticality.
- Mercury manometers are being withdrawn from Indian plants and labs on account of mercury-handling restrictions; electronic manometers and digital pressure calibrators have replaced them.
Elastic (mechanical) elements
- Principle: pressure deforms an elastic element within its proportional limit; deflection is amplified mechanically and shown on a dial, or converted to an electrical signal.
- Bourdon tube: a flattened (elliptical/oval) tube bent into a C; internal pressure tries to make the section circular, uncoiling the tube and moving the tip, which drives a link-lever-pinion and gear train against a hairspring. C-type covers roughly 0.6 bar up to several hundred bar (special designs to ~1000 bar+). Spiral and helical forms give more tip travel for the same stress - helical types suit high pressure and direct pen drive in recorders. Twisted tube used in torque-balance transmitters. Materials: phosphor bronze, beryllium copper, alloy steel, SS 316L/Monel for corrosive service.
- Diaphragm: a thin circular metal (or elastomer) plate, flat or corrugated, clamped at the edge; centre deflection ∝ pressure for small deflection. Good for low pressures, corrosive and viscous fluids, small hysteresis, high natural frequency. A capsule is two diaphragms welded at the rim (double deflection); stacked capsules give still more travel.
- Bellows: a thin seamless convoluted tube; large deflection for low pressure, usually opposed by a calibrated spring (so spring, not metal, sets the range). Used in 0-few bar pressure switches, differential units and absolute elements (one bellows evacuated).
- Advantages: rugged, cheap, needs no power, directly readable. Limitations: hysteresis and fatigue, set drift, temperature error from change in elastic modulus, mechanical wear, poor dynamic response, non-linear at large deflection.
- Dial gauge practice: select range so the normal reading sits in the middle of the span - normal operating pressure between about 25% and 75% of span, ideally near 50%; accuracy classes of dial gauges and their test methods are covered by the Indian Standard for pressure and vacuum gauges (IS 3624) - verify the current edition and class designations before quoting figures.
Electrical and electronic pressure transducers
- Strain-gauge type: bonded foil or thin-film gauges on a diaphragm, cylinder or Bourdon tube in a four-arm Wheatstone bridge (two gauges in tension, two in compression) - gives maximum output and inherent temperature compensation. Output in mV/V; needs excitation and amplification.
- Piezoresistive / MEMS: gauges diffused or implanted into a silicon diaphragm; gauge factor 50-200 (far above metal's ~2), so output is large, size tiny and cost low - but strongly temperature-dependent, hence on-chip compensation. Basis of almost all modern low-cost sensors (automotive MAP sensors, mobile barometers, BP monitors).
- Capacitive: diaphragm forms one plate of a capacitor; deflection changes the gap, so C = εA/d changes. Differential (three-plate) construction with oil-filled isolating diaphragms gives excellent accuracy (0.05-0.075% of span), high stability and high turndown - the standard DP transmitter sensor.
- Inductive / LVDT type: diaphragm or bellows moves an LVDT core or changes reluctance; robust, good output, but larger and speed-limited by the carrier frequency.
- Piezoelectric: quartz, tourmaline or PZT generates charge Q = d·F when stressed. Dynamic pressures only (charge leaks away), so used for blast, engine cylinder pressure, shock, cavitation; needs a charge amplifier; very high natural frequency (>100 kHz).
- Resonant / vibrating-wire and resonant silicon: pressure changes tension or stiffness and hence the resonant frequency - frequency output, very stable, used in high-accuracy and subsea instruments.
- Optical / fibre-Bragg-grating sensors: immune to EMI, suited to high-temperature and intrinsically safe applications.
- Smart transmitters: sensor + ADC + microprocessor + DAC giving 4-20 mA two-wire output with HART digital overlay, or FOUNDATION Fieldbus/PROFIBUS-PA, or wireless (WirelessHART/ISA100). Features: characterisation and linearisation, temperature and static-pressure compensation, remote re-ranging, damping, square-root extraction, diagnostics (plugged-impulse-line detection), 3.6/21+ mA failure signalling.
- Pressure switches: diaphragm/bellows against a set spring operating a microswitch, with adjustable set point and dead band - used for alarms, pump cut-in/cut-out and interlocks.
Vacuum and very high pressure
- Low vacuum (atmosphere to ~1 mbar): Bourdon and diaphragm/capsule vacuum gauges, U-tube manometer.
- McLeod gauge: compresses a known volume V of gas into a capillary of area a by raising mercury; from Boyle's law p ≈ a·h²/V. Primary standard for about 10⁻¹ to 10⁻⁵ mmHg; manual, slow, does not work for condensable vapours, and mercury-based.
- Thermal conductivity gauges: gas thermal conductivity falls with pressure in the Knudsen region. Pirani gauge - a heated filament in a Wheatstone bridge; range ~10⁻⁴ to 1 mbar; thermocouple gauge - similar, with a thermocouple reading filament temperature. Both are gas-species dependent.
- Ionization gauges: gas is ionised and ion current ∝ pressure. Hot-cathode (Bayard-Alpert) for ~10⁻³ down to 10⁻¹⁰ mbar; cold-cathode (Penning) more rugged, slightly narrower range.
- Knudsen gauge (molecular momentum) and spinning-rotor gauge for high-vacuum transfer standards.
- High pressure: bulk-modulus (pressure cell) gauges, electrical resistance pressure gauge (Bridgman gauge) - manganin wire resistance rises with pressure, usable to ~10 kbar; thick-walled strain-gauged cylinders.
- Dead-weight (piston) gauge tester: calibrated masses on a piston of known effective area in a fluid-filled cylinder; p = Mg/A_eff with corrections for local gravity, air buoyancy, temperature (area expansion) and head. The primary pressure standard in Indian calibration labs, traceable to CSIR-NPL.
Installation, errors and dynamic behaviour
- Impulse lines: slope at least 1:10 so gas lines drain and liquid lines self-vent; keep short; use an isolation/root valve plus a bleed; a manifold (3-valve or 5-valve) on DP transmitters for zeroing without over-ranging.
- Head (elevation) error: a liquid-filled line between tapping and instrument adds ±ρgh - corrected by zero adjustment, not by ignoring it.
- Steam service: a siphon (pigtail) loop keeps condensate between steam and the Bourdon tube. Pulsating service: a snubber or needle restrictor; corrosive/slurry/viscous/hygienic service: a diaphragm seal (chemical seal) with capillary and fill fluid, or a flush-mounted sanitary diaphragm.
- Wet leg / dry leg on a closed vessel, and capillary fill-fluid temperature effects, are the usual sources of DP zero shift.
- Other errors: ambient and process temperature coefficients, static-pressure (line-pressure) span error on DP cells, over-range, vibration, mounting position, EMI on analogue loops, air trapped in liquid lines, plugged taps.
- Dynamic response: a pressure line plus instrument volume behaves as a second-order system (acoustic inertance with line resistance and element compliance); narrow, long lines and trapped gas lower the natural frequency and make the response sluggish. For fast transients use a flush-mounted piezoelectric or piezoresistive sensor with negligible line length.
- Calibration: rising and falling (hysteresis) checks at 0, 25, 50, 75, 100% of span against a dead-weight tester or a NABL-traceable digital calibrator; record as-found and as-left; certificate under ISO/IEC 17025. Field work uses a hand pump/pneumatic calibrator with a loop calibrator for the 4-20 mA output.
- Indian codes and practice: Indian Boiler Regulations require a pressure gauge with a test cock and a graduated dial on every boiler; pressure vessels and gas cylinders fall under PESO (Gas Cylinders Rules, Static and Mobile Pressure Vessels Rules) with certified gauges and relief devices; safety-relief and gauge requirements for unfired pressure vessels follow IS 2825 / ASME BPVC Section VIII; gauges in hazardous areas must be certified to the IS/IEC 60079 series (Ex ia, Ex d) with enclosure protection to IS/IEC 60529; instruments used in trade come under the Legal Metrology Act, 2009; workplace safety obligations now sit under the Occupational Safety, Health and Working Conditions Code, 2020, in force since 21 November 2025, which replaced the Factories Act.
Key points
- Pressure - normal force per unit area; SI unit pascal; 1 bar = 10⁵ Pa = 100 kPa ≈ 1.0197 kgf/cm² ≈ 14.5 psi.
- Absolute = gauge + atmospheric - a gauge reading zero means the fluid is at atmospheric pressure, not at vacuum.
- Manometer - primary device, p = ρgh; accuracy depends only on density and length measurement.
- Inclined tube - sensitivity gain 1/sinθ; the standard low-range draft gauge.
- Two-liquid micromanometer - magnification from (ρ₁ − ρ₂); for pressures of a few pascals.
- Bourdon tube - oval tube tries to become circular and uncoils; C, spiral, helical and twisted forms; wide range, no power needed, has hysteresis.
- Diaphragm/capsule/bellows - low-pressure elastic elements; bellows range is set by the opposing spring.
- Strain gauge bridge - four active arms give maximum output and temperature compensation; output in mV/V.
- Piezoresistive silicon - gauge factor 50-200 versus ~2 for metal foil, but needs temperature compensation.
- Capacitive DP cell - highest industrial accuracy (≈0.05% of span) and turndown; heart of the DP transmitter.
- Piezoelectric - dynamic pressure only, charge output, very high frequency response; useless for steady pressure.
- Dead-weight tester - p = Mg/A_eff; the workshop primary standard, traceable to CSIR-NPL.
- McLeod gauge - Boyle's-law compression, p ≈ ah²/V, 10⁻¹ to 10⁻⁵ mmHg; primary vacuum standard.
- Pirani/thermocouple gauge - thermal conductivity, ~10⁻⁴ to 1 mbar, gas dependent; ionization gauge down to 10⁻¹⁰ mbar.
- 4-20 mA two-wire - live zero detects open circuit; p = (I − 4)/16 × span + LRV.
- Siphon for steam, snubber for pulsation, diaphragm seal for corrosive or clogging fluids.
- Head correction - ρgh of the filled impulse line must be zeroed out, not ignored.
- Select range so normal operating pressure falls near mid-span, and never read a gauge near its over-range limit.
Formulas and examples
U-tube differential: p₁ − p₂ = (ρ_m − ρ_f) g h [ρ_m = manometer liquid, ρ_f = fluid above it]
Inclined manometer: p = ρ g L sinθ; sensitivity = L/h = 1/sinθ
Two-liquid micromanometer: Δp = (ρ₁ − ρ₂) g h (reservoir-area correction neglected)
Absolute = gauge + atmospheric; vacuum (abs) = p_atm − vacuum reading
Flat clamped circular diaphragm, centre deflection y = 3 p R⁴ (1 − ν²) / (16 E t³) [R = radius, t = thickness, E = Young's modulus, ν = Poisson's ratio]; keep y < t/3 for linearity
Strain-gauge full bridge: V_o = V_ex · GF · ε (four active gauges); quarter bridge: V_o = V_ex · GF · ε / 4
Capacitive sensor: C = ε₀ε_r A / d; differential output ∝ (C₁ − C₂)/(C₁ + C₂)
Piezoelectric: Q = d·F = d·p·A; open-circuit voltage V = Q/C
Dead-weight tester: p = M g / A_eff
McLeod gauge: p = a h² / (V − a h) ≈ a h² / V
Current-loop conversion: p = (I − 4 mA)/(16 mA) × span + LRV; V across R = I·R
Unit conversions: 1 atm = 101.325 kPa = 760 mmHg = 1.01325 bar = 10.33 mH₂O; 1 torr = 133.322 Pa; 1 kgf/cm² = 98.0665 kPa; 1 mmWC = 9.81 Pa
Example 1 - mercury manometer, gauge and absolute. A U-tube mercury manometer connected to a gas line reads h = 250 mm. ρ_Hg = 13,600 kg/m³, g = 9.81 m/s², barometer 101.325 kPa.
- Gauge pressure = 13,600 × 9.81 × 0.250 = 33,354 Pa = 33.35 kPa = 0.334 bar(g).
- Absolute = 33.35 + 101.325 = 134.68 kPa(a).
- In kgf/cm² = 33.354/98.0665 = 0.340 kgf/cm²(g).
Example 2 - inclined draft gauge. Tube at 30° to horizontal, liquid SG 0.8, reading along the tube L = 150 mm.
- Vertical head h = 0.150 × sin30° = 0.075 m; p = 800 × 9.81 × 0.075 = 588.6 Pa = 60 mmWC.
- A vertical tube of the same liquid would have shown only 75 mm, so sensitivity gain = 1/sin30° = 2.
Example 3 - dead-weight tester. Piston effective area A_eff = 1.000 cm² = 1.000 × 10⁻⁴ m²; total mass loaded (piston + weights) = 20.00 kg; local g = 9.790 m/s².
- F = 20.00 × 9.790 = 195.8 N; p = 195.8 / 1.000 × 10⁻⁴ = 1.958 × 10⁶ Pa = 19.58 bar(g).
- If standard g = 9.80665 had been assumed, p would read 19.61 bar - an error of about 0.17%, which is why local gravity is applied in NABL-traceable certificates.
Example 4 - diaphragm sizing. Stainless-steel flat diaphragm, R = 10 mm, t = 0.3 mm, E = 200 GPa, ν = 0.3, p = 1 bar = 10⁵ Pa.
- Numerator = 3 × 10⁵ × (0.01)⁴ × (1 − 0.09) = 3 × 10⁵ × 10⁻⁸ × 0.91 = 2.73 × 10⁻³.
- Denominator = 16 × 2 × 10¹¹ × (3 × 10⁻⁴)³ = 3.2 × 10¹² × 2.7 × 10⁻¹¹ = 86.4.
- y = 2.73 × 10⁻³ / 86.4 = 3.16 × 10⁻⁵ m = 0.0316 mm, i.e. about 0.105 t - safely in the linear region.
Example 5 - 4-20 mA transmitter and piezoelectric sensor.
- A transmitter is ranged 0-25 bar(g) and the loop current is 13.6 mA. p = (13.6 − 4)/16 × 25 = (9.6/16) × 25 = 15.0 bar(g). Across a 250 Ω sense resistor the DCS reads 13.6 mA × 250 Ω = 3.40 V (of the 1-5 V span).
- A quartz sensor has charge sensitivity d = 25 pC/N, diaphragm area 1 cm², internal capacitance 500 pF. A 50 bar transient gives F = 50 × 10⁵ × 10⁻⁴ = 500 N; Q = 25 × 10⁻¹² × 500 = 1.25 × 10⁻⁸ C = 12.5 nC; V = 12.5 nC / 500 pF = 25 V open circuit.
Example 6 - vacuum conversions. A gauge shows 600 mmHg vacuum with barometer 760 mmHg.
- Absolute = 760 − 600 = 160 mmHg = 160/760 × 101.325 = 21.33 kPa(a) = 0.213 bar(a).
- A McLeod gauge with V = 100 cm³ (= 10⁵ mm³) and capillary area a = 1 mm² reading h = 10 mm gives p ≈ 1 × 10² / 10⁵ = 1 × 10⁻³ mmHg.
| Device | Typical range | Accuracy (typical) | Notes |
|---|---|---|---|
| U-tube / inclined manometer | 0-5 kPa (to ~2 bar with Hg) | 0.1-1% | Primary, no power, bulky |
| Bourdon dial gauge | 0.6 bar to 700 bar | 0.5-2.5% of span | Local indication, hysteresis |
| Diaphragm / capsule | 0-0.1 bar to 10 bar | 0.5-1% | Low pressure, corrosive service |
| Bellows | 0-0.05 to 5 bar | 0.5-1% | Switches, DP units |
| Strain gauge / piezoresistive | 0-0.1 bar to 1000 bar | 0.1-0.5% | Small, fast, temp-compensated |
| Capacitive DP transmitter | 0-0.1 kPa to 20 MPa | 0.05-0.1% of span | High turndown, HART |
| Piezoelectric | to 10 kbar, dynamic only | 1% | kHz response, no static reading |
| Pirani gauge | 10⁻⁴ to 1 mbar | 10-20% | Gas dependent |
| Ionization gauge | 10⁻³ to 10⁻¹⁰ mbar | 10-30% | High vacuum only |
| Dead-weight tester | 0.2 bar to 1000+ bar | 0.015-0.05% | Calibration standard |
Where it is used
- Refineries, petrochemicals and fertiliser plants (IOCL, BPCL, RIL, GAIL, IFFCO): DP transmitters on orifice flow and vessel level, pressure transmitters on compressors and columns, Ex-certified instruments with diaphragm seals on fouling services.
- Power and boiler plants (NTPC, BHEL, Thermax, Forbes Marshall): drum pressure and furnace draft under Indian Boiler Regulations, steam-line gauges with siphons, pressure switches for burner interlocks.
- Automotive and mobility (Bosch, Maruti Suzuki, Tata Motors, TVS): MEMS MAP and barometric sensors, common-rail fuel pressure (above 1,500 bar), tyre-pressure monitoring, brake-circuit pressure.
- Pharma, food and water utilities: sanitary flush-diaphragm transmitters, filter-differential monitoring, pump discharge and network pressure logging in city water SCADA under AMRUT/Jal Jeevan-type projects.
- Calibration and quality roles: NABL-accredited calibration labs, dead-weight testers and digital calibrators traceable to CSIR-NPL; PESO-related certification of cylinders and SMPV installations.
- Job roles: instrumentation engineer, control-and-instrumentation (C&I) engineer, commissioning and loop-checking engineer, calibration technician, product/application engineer with Emerson, Yokogawa, ABB, Honeywell, Siemens, Endress+Hauser, Wika, and panel/field roles with L&T and Tata Projects.
Interview questions
- Q: What is the difference between gauge and absolute pressure? A: Gauge pressure is measured relative to local atmospheric pressure, absolute relative to perfect vacuum; absolute = gauge + atmospheric, and an ordinary gauge reads zero at atmosphere because its case interior is open to air.
- Q: How does a Bourdon tube work? A: An oval-section tube bent into an arc tends to become circular under internal pressure, which uncoils it; the tip motion is amplified by a link, lever and gear train to move the pointer.
- Q: Why would you choose a diaphragm element over a Bourdon tube? A: For low pressures, corrosive, viscous or slurry fluids, hygienic service, better linearity and higher natural frequency; a Bourdon tube has a narrow bore that clogs and has more hysteresis.
- Q: Why can a piezoelectric sensor not measure static pressure? A: The generated charge leaks away through the crystal and amplifier input resistance, so the output decays; it responds only to changing (dynamic) pressure.
- Q: Why is 4-20 mA used instead of 0-20 mA or a voltage signal? A: The 4 mA live zero distinguishes a true zero reading from a broken wire or dead transmitter, a current signal is immune to line-resistance voltage drop, and the 4 mA also powers a two-wire transmitter.
- Q: What is turndown and why do capacitive DP cells have high turndown? A: Turndown is maximum span divided by the smallest calibrated span; a capacitive cell's low noise, high stability and digital linearisation let it be re-ranged to a small fraction of maximum span while keeping accuracy.
- Q: What is a dead-weight tester and what corrections does it need? A: A piston-cylinder of known effective area loaded with calibrated masses, giving p = Mg/A_eff; it needs corrections for local gravity, air buoyancy, temperature expansion of the area, fluid head and piston friction (hence rotation).
- Q: Why use a siphon on a steam pressure gauge? A: The pigtail loop traps condensate so the hot steam never reaches the Bourdon tube, protecting the element from thermal damage and zero drift.
- Q: How would you measure pressure in a system with heavy pulsation? A: Fit a snubber or needle-valve restrictor, use a liquid-filled (glycerine) gauge for damping, or apply electronic damping in a smart transmitter - and size the range for the peak, not the average.
- Q: Why is a McLeod gauge unsuitable for condensable vapours? A: It relies on Boyle's law during compression; a condensable vapour liquefies as it is compressed, so the measured capillary height understates the true pressure.
- Q: How does a Pirani gauge measure vacuum, and what is its limitation? A: Below about 1 mbar the thermal conductivity of a gas falls with pressure, so the temperature (and resistance) of a heated filament in a bridge indicates pressure; its calibration depends on the gas species and it fails at high vacuum where conduction is negligible.
- Q: A DP transmitter on an orifice reads zero flow while the plant is running. What do you check? A: Equalising valve left open on the manifold, both isolation valves shut, plugged or air-locked impulse lines, reversed high/low connections, or an orifice plate installed backwards.
- Q: Why is static-pressure error specified separately for a DP transmitter? A: High line pressure distorts the sensor body and shifts zero and span even when the differential is unchanged, so the specification gives a per-bar error that is zeroed out at operating pressure.
Common mistakes
- Adding or omitting atmospheric pressure wrongly - mixing gauge and absolute values in the same calculation, especially in gas-law and vacuum problems.
- Using the manometer liquid density alone in a differential manometer instead of (ρ_m − ρ_f), and forgetting that g and liquid density vary with location and temperature.
- Forgetting the sinθ factor on an inclined manometer, or treating the long tube reading as the head itself.
- Ignoring the impulse-line head when the transmitter is mounted above or below the tapping, then "fixing" the offset by re-spanning instead of re-zeroing.
- Using a quarter-bridge formula (V_ex·GF·ε/4) where four active gauges are fitted, or vice versa, and mixing µε with strain in the same expression.
- Expecting a piezoelectric sensor to hold a static reading, or a Bourdon gauge to follow a millisecond transient.
- Unit slips - kgf/cm² treated as bar, mmWC as mmHg, bar as Pa - and selecting a gauge whose normal reading sits in the bottom 10% or above 90% of span.
- Opening a 3-valve manifold in the wrong sequence and over-ranging the DP cell, or calibrating only upward and never checking hysteresis on the downward pass.
Related topics
Transducers and sensors basics · Flow measurement · Level measurement · Temperature measurement · Signal conditioning and bridge circuits · Process control and final control elements